HIGH VOLTAGE BREAKDOWN TESTER

MODEL : HV-3KV DC

Specifications

High Voltage Output
Output Voltage : 0 to 3 KV DC Hz Setting Variable using Autotransformer on Front Panel
Readout : Analogue meter with scale length 90 mm
Resolution : 100 Volts
Accuracy : ±5% of FSD
Leakage Current Measurement
Readout : Analogue meter with scale length 90 mm, 15mA FSD
Resolution : 300 Micro amp
Accuracy : ±5% of FSD
Trip Circuit : Trip current 1mA to 15 mA adjustable
Trip Annunciation : Trip indication annunciation on LED Lamp and Buzzer
Test Leads : Test leads will be fully insulated alligator clips to convenient for testing most small devices/components like relay
Operating Conditions
Operating Power : 200 – 250 volts, 50 Hz Power consumption 180 VA
Operating Environment : 0 to 55 degree centigrade relative humidity < 90% Non-condensing
Cabinet
Mild Steel Structure with Aluminium Front Panel and Back Panel Powder Coated with Siemens Gray colour and Top and side cover with DA Gray colour
Dimension : 340mm width X 350mm depth X 280mm Height
Weight : 13.5 Kg.